|
Electronic Transmitter Adjusted Range
This figure shows how the range of the measured variable, on the left, relates to the
range of the transmitted signal, on the right. The measured variable is also the input and
the transmitted signal is the output. The measurement range is 100 to 500 degrees C. Thus,
the span of measurement is 400 degrees C. The transmitter range is 4 to 20 milliAmps (mA),
and the transmitter span is 16 mA. At the lowest measurement (100 C), the transmitter
output should be 4 mA. At fifty percent of the measurement (300 C), the transmitter output
should be 12 mA.
Using standard signal ranges, current output in a process can be readily calculated
using the formula for steady-state gain:
Steady-State Gain (K) = Change in Output / Change in Input that caused
output change
The gain for this transmitter is .04 (for example, 16 divided by 400). Once the gain is
known, the temperature can be calculated from the output of the transmitter. For example,
let's say the the transmitter output is 8 mA. Remember, 4 mA is zero percent of span, so
(subtracting 4 mA from 8 mA) the transmitter output change is 4 mA. Dividing 4 mA by the
gain (.04) produces an answer of 100. The lowest temperature in the range is 100 degrees
C. Adding 100 and 100 produces the actual temperature 200 degrees C.
|